TECNAI T12 September 2013 Start-of-Session Steps 1. Add LN 2to ACD dewar. Check level during session. DO NOT allow to dry! 2. Log into Windows with your username and password. 3. Open TECNAI USER INTERFACE; Open TIA. 4. Load Sample. Put sample onto chosen sample holder (single tilt; double tilt.)

How to align the beam on a tecnai?

Select and align Condenser Aperture. Condense beam to cross-over with INTENSITY knob. Center beam with Track Ball. Spread beam with INTENSITY until beam is same size as TEM screen. Align Condenser Aperture so beam edge matches screen edge. (See Step 13, Objective Aperture Alignment, for image showing location of aperture alignment controls.) 9.

What’s the difference between tiaprogram and tecnai?

The TIAprogram displays on the right-hand monitor, while Tecnai User Interface(which controls the TEM) displays on the left- hand monitor. The mouse cursor can move freely between the right-hand and left-hand monitors.

What are the start of session steps for tecnai?

Start-of-Session Steps 1. Add LN 2to ACD dewar. Check level during session. DO NOT allow to dry! 2. Log into Windows with your username and password. 3. Open TECNAI USER INTERFACE; Open TIA. 4. Load Sample. Put sample onto chosen sample holder (single tilt; double tilt.) Observe O-ring under optical microscope and remove dirt and dust.